The Xhibit Central Station (model 96102) gives clinicians central monitoring of data from networked Spacelabs patient monitors and telemetry transmitters. It also generates visible and audible alarms. A four-way DisplayPort splitter copies the station's display and audio to as many as three remote displays, for example in corridors, care alcoves or break rooms. FDA's recall entry describes the splitter as a third-party external accessory. It gives the manufacturer's reason for recall as an "off-the-shelf display port component that has audio/video failures" and lists the FDA-determined cause as "Unknown/Undetermined by firm". According to Spacelabs' field safety notice, the problem was found through a review of product complaints about lost video and audio. The notice says splitters "may fail to perform as intended and result in degraded or complete loss of alarm audio and video" and that unqualified splitters may pose the same risks. It states that clinical staff may then be unable to detect an alarm condition and may fail to respond to a medical condition that requires intervention, and that serious injury or death is foreseeable.
On 4 June 2026 Spacelabs emailed customers an urgent medical device correction letter. The letter told them to stop using splitter 010-1975-00 or any unapproved splitter, notify staff, acknowledge the notice, dispose of the affected components, and install a newly qualified replacement (model 010-1977-00, kit 020-0235-00). According to the French-language notice published by ANSM, customers who bought the affected splitter from Spacelabs receive the replacement free of charge. The English notice published by the Dutch inspectorate IGJ sets 30 August 2026 as the completion date. FDA lists 6,074 units in commerce, US nationwide distribution and several other countries. It posted the recall on 7 August 2026, and the recall remained open.[1,2,3]