Incident

Luna G3 BPAP devices record the set pressure instead of the pressure delivered when Targeted VT is on; Class I recall

FDA classified a recall of the React Health (3B Medical) Luna G3 BPAP System, model LG3800-30VT, as Class 1. With Targeted VT enabled, the device records and transmits the configured inspiratory pressure instead of the pressure actually delivered. React Health says this could lead clinicians to adjust settings based on inaccurate data, which could delay therapy.

Sources checked when written 28 November 2026

TierThe sources document a fault in recorded and transmitted pressure data while the devices kept delivering therapy. They include instructions to switch off Targeted VT and swap devices, but do not document that care actually changed; wrong output does not change tier. How tiers are assigned.

What happened

On 12 November 2026 FDA posted a Class 1 recall (Z-0409-2027) of the Luna G3 BPAP System, model LG3800-30VT. The FDA record describes it as a bi-level positive airway pressure device for treating adult obstructive sleep apnea. The recalling firm is 3B Medical, Inc. (React Health). The FDA record gives the reason as "The configured IPAP value is recorded and transmitted instead of the actual delivered inspiratory pressure when Targeted VT is enabled", and FDA determined the cause as device design. React Health's notice says that with Targeted VT enabled, affected devices do not record the pressure actually delivered on the internal SD card. They also do not transmit it to the company's cloud-based management system, React Health Connect. Devices without Targeted VT enabled are not affected. According to React Health, the inaccurate pressure data in historical therapy reports "could lead a clinician to adjust device settings based on inaccurate therapy pressure data, which could result in delayed therapy until the setting is corrected." The FDA record dates the firm's initiation to 8 October 2026 and says the firm began notifying customers on 9 October. React Health dates its original notification 8 October and published an update on 17 November. The two sources give different counts. The FDA record lists 12,672 units in commerce in the US, including Puerto Rico. React Health's notice says 7,533 devices were distributed in the US and Puerto Rico. The FDA record tells customers to: - discontinue Targeted VT immediately, - remove affected devices from clinical service, - use unaffected PAP devices for patient support, - install a firmware correction from React Health. React Health's notice tells patients to contact their provider or equipment supplier. It also tells them not to change settings or stop therapy on their own, and to keep using the device as prescribed until they receive guidance. It names firmware G3-2.00.80 as the correction. It says data collected with Targeted VT enabled before the upgrade will not be changed and will still show the previously recorded pressures.[1,2]

Documented harm

React Health says it has received fifteen complaints, including two serious adverse reports, associated with the issue. Its notice does not describe them or say the fault caused them, and the FDA record reports no injuries or deaths.

What it teaches

  • When a device adjusts therapy itself, remote reports must show what it delivered, not what it was set to, or clinicians reviewing them cannot see what the automation did.
  • Data paths used only for review, such as SD-card logs and cloud dashboards, need the same verification as the therapy path, because clinicians change settings based on them.
  • A firmware fix that corrects data going forward leaves the historical record unchanged, so earlier reports still need to be flagged when they are interpreted.

Sources

  1. Class 1 Device Recall Luna G3 BPAP System (Z-0409-2027). U.S. Food and Drug Administration, 12 November 2026. Primary Dataset · link checked 2026-11-23
  2. Medical Device Correction Update: Luna G3 BPAP System (Model LG3800-30VT) Voluntary Recall – Targeted Pressure Data Reporting Correction. React Health (3B Medical, Inc.), 17 November 2026. Supporting Guidance · link checked 2026-11-23

LATITUDE Clarity server software can hold back AF Burden alerts from LUX-Dx implantable cardiac monitors

Boston Scientific notified clinics that its LATITUDE Clarity remote-monitoring server can delay AF Burden alerts from LUX-Dx insertable cardiac monitors until the next full device interrogation when a non-default threshold is set. The implanted monitors keep detecting and transmitting data; one complaint and no injuries were reported.

Philips PIC iX central monitoring stations can freeze when the watchdog recovery fails; central alarms may be lost

Philips issued a correction for Patient Information Center iX versions 4.0–4.5 after receiving reports that some systems froze and could not be used. According to Health Canada and Philips' notice, a failed watchdog recovery can leave the system in an unknown state in which central monitoring and alarms may not work.

CrowdStrike Falcon content update crashes Windows hosts, including hospital systems

A faulty Rapid Response Content update to CrowdStrike's Falcon sensor crashed about 8.5 million Windows devices worldwide. Outside-in measurement found disrupted services at 759 of 2,232 US hospitals studied.

PathDevices → Connectivity & data → Human handoff

Study of 2017-2025 FDA data: neurosurgical software recalls, mostly Class I, with adverse-event reports continuing after recall

A peer-reviewed analysis in Neurosurgical Focus identified 20 software-associated neurological device recalls initiated in the US from 2017 to 2025, more than half of them FDA Class I. It counted 16,333 adverse-event notifications for these devices, more than 10,000 of them reported after formal recall notification for two devices.

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

All incidents · Devices · Human handoff

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.