Incident

Tandem Mobi insulin pumps: false occlusion alarms suspend insulin delivery; Class 1 correction

Tandem Diabetes Care began a correction of Tandem Mobi insulin pumps in September 2026 because the pumps can raise false-positive occlusion alarms that suspend insulin delivery; FDA lists it as a Class 1 recall. A company spokesperson reported 230 adverse events requiring medical intervention or hospitalization as of 29 October 2026, and no deaths.

Sources checked when written 28 November 2026

TierThe pumps stay in use and the failure is a false alarm (wrong output), which under the method does not change tier. How tiers are assigned.

What happened

FDA's recall database lists recall Z-0316-2027, headed "Class 1 Device Recall Tandem Mobi", initiated by the firm on 23 September 2026, posted on 30 October 2026, and open. According to the FDA record, pumps may give false-positive occlusion alarms when no insulin flow obstruction is present and suspend insulin delivery; MedTech Dive reports that delivery stays suspended until the user acknowledges the alarm. According to MPO and MedTech Dive, Tandem found that increased friction in the drive module can reduce available motor power and, under certain conditions, make motor speed fall below the occlusion detection threshold. MedTech Dive, citing the FDA database entry, reports that the recall covers more than 80,900 devices, most distributed in the US, and that FDA said repeat false alarms can increase the risk of high blood glucose. The FDA record lists serial numbers of pumps with frequent false-positive occlusion alarms to be exchanged. Users were told they may keep using their pumps but must respond to every occlusion alarm, follow troubleshooting instructions and keep a backup method of insulin delivery. Users should contact customer support if alarms continue (MedTech Dive, MPO). MPO reports that Tandem contacted users whose pumps were identified for replacement and is updating its testing procedures to better identify affected devices. This action is separate from Tandem's earlier Mobi correction, classified on 5 November 2025, for a false motor-failure alert (Malfunction 12) in software versions 7.6.0.1, 7.6.0.3 and 7.7.0.1, which FDA said had four reported serious injuries and no deaths as of 4 November 2025.[1,2,3,4,5]

Documented harm

FDA's recall notice says that, as of 28 October 2026, Tandem Diabetes had reported 1,248 serious injuries and no deaths associated with the issue. A Tandem spokesperson gave MedTech Dive a lower figure, 230 adverse events requiring medical intervention or hospitalization as of 29 October 2026, and MPO reported 63 such events as of 10 September 2026 without naming the source of that count. The sources do not explain the difference.

What it teaches

  • Detection thresholds tied to mechanical margins can drift as parts wear or vary, turning a safety alarm into a cause of interrupted therapy.
  • A false-positive alarm that stops therapy carries its own clinical risk, so false-alarm rates need the same scrutiny as missed detections.
  • Patients on automated delivery devices need a backup method that is ready to use, not just prescribed.

Sources

  1. Class 1 Device Recall Tandem Mobi (Z-0316-2027). U.S. Food and Drug Administration, 30 October 2026. Primary Dataset · link checked 2026-11-14
  2. Tandem recalls Mobi insulin pump, reports 230 adverse events. MedTech Dive, 11 November 2026. Supporting Journalism · link checked 2026-11-14
  3. Tandem Corrects U.S. Mobi Insulin Pumps Over Occlusion Alarms. Medical Product Outsourcing (MPO), 9 November 2026. Supporting Journalism · link checked 2026-11-14
  4. Insulin Pump Correction: Tandem Diabetes Care Issues Correction for Tandem Mobi Insulin Pumps. U.S. Food and Drug Administration, 5 November 2025. Primary Official report · link checked 2026-11-14
  5. Insulin Pump Recall: Tandem Diabetes Removes Tandem Mobi Insulin Pumps. U.S. Food and Drug Administration, 25 November 2026. Primary Regulation · link checked 2026-11-26

Study of 2017-2025 FDA data: neurosurgical software recalls, mostly Class I, with adverse-event reports continuing after recall

A peer-reviewed analysis in Neurosurgical Focus identified 20 software-associated neurological device recalls initiated in the US from 2017 to 2025, more than half of them FDA Class I. It counted 16,333 adverse-event notifications for these devices, more than 10,000 of them reported after formal recall notification for two devices.

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

Luna G3 BPAP devices record the set pressure instead of the pressure delivered when Targeted VT is on; Class I recall

FDA classified a recall of the React Health (3B Medical) Luna G3 BPAP System, model LG3800-30VT, as Class 1. With Targeted VT enabled, the device records and transmits the configured inspiratory pressure instead of the pressure actually delivered. React Health says this could lead clinicians to adjust settings based on inaccurate data, which could delay therapy.

VOCSN ventilators: loose internal flow-transducer connector can cause reboots; FDA early alert and Class I recall

React Health (Ventec Life Systems) warned that the internal flow transducer cable connector in VOCSN ventilators can unseat and cause reboots and recurring System Fault 10 alarms, which could interrupt ventilation. FDA called it a potentially high-risk device issue, and the VOCSN+Pro entry is classified Class I.

Philips Lumify ultrasound demands online re-registration and blocks scanning in pre-hospital settings

FDA issued an Early Alert on 9 October 2026 and on 19 October 2026 said the issue had been classified as a Class I recall. Philips had reported that its Lumify ultrasound app may stop recognising a registered transducer and block scanning until the transducer is re-registered, which requires Wi-Fi or cellular access. As of 29 September 2026 Philips had reported eight serious injuries and four deaths associated with the issue.

PathConnectivity & data → Devices

All incidents · Devices

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.

Dealing with an incident right now? This site is a reference, not an incident-response service. Activate your organisation's emergency operations plan and incident command, and: