FDA's recall database lists recall Z-0316-2027, headed "Class 1 Device Recall Tandem Mobi", initiated by the firm on 23 September 2026, posted on 30 October 2026, and open. According to the FDA record, pumps may give false-positive occlusion alarms when no insulin flow obstruction is present and suspend insulin delivery; MedTech Dive reports that delivery stays suspended until the user acknowledges the alarm. According to MPO and MedTech Dive, Tandem found that increased friction in the drive module can reduce available motor power and, under certain conditions, make motor speed fall below the occlusion detection threshold. MedTech Dive, citing the FDA database entry, reports that the recall covers more than 80,900 devices, most distributed in the US, and that FDA said repeat false alarms can increase the risk of high blood glucose. The FDA record lists serial numbers of pumps with frequent false-positive occlusion alarms to be exchanged.
Users were told they may keep using their pumps but must respond to every occlusion alarm, follow troubleshooting instructions and keep a backup method of insulin delivery. Users should contact customer support if alarms continue (MedTech Dive, MPO). MPO reports that Tandem contacted users whose pumps were identified for replacement and is updating its testing procedures to better identify affected devices. This action is separate from Tandem's earlier Mobi correction, classified on 5 November 2025, for a false motor-failure alert (Malfunction 12) in software versions 7.6.0.1, 7.6.0.3 and 7.7.0.1, which FDA said had four reported serious injuries and no deaths as of 4 November 2025.[1,2,3,4,5]