Incident

Baxter Novum IQ infusion pumps: underinfusion and software faults lead to Class I corrections and a distribution hold

FDA issued early alerts in July and August 2025 about Baxter's Novum IQ large volume pump, which can underinfuse after certain rate increases, and about software faults in both Novum IQ pumps. The corrections were classified as Class I recalls. Baxter states that it temporarily stopped distributing and installing the large volume pump in the US and Canada in July 2025, except in cases of medical necessity.

Sources checked when written 28 November 2026

TierThe sources document wrong or interrupted delivery, bedside workarounds and replacement with other pumps, but not care moving off automated infusion. How tiers are assigned.

What happened

On 22 July 2025 FDA issued an early alert about Baxter's Novum IQ large volume pump (LVP). According to the FDA notice, Baxter reported that the pump can underinfuse when it moves from one flow rate to a rate more than double the first, for example during a rate change or bolus, and that in the worst case no delivery may occur. Baxter also reported more customer complaints of over- and underinfusion, possibly caused by tubing misloaded into the pump channel. Baxter's 14 July letter told customers to change the administration set, or switch to another pump with a new set, at the first safe opportunity. Where that delay was unacceptable, staff were to move the set about 0.5 inches downstream before starting a bolus or a rate change above 100%. FDA classified this correction as a Class I recall on 22 December 2025. A second correction, in a Baxter letter dated 4 August 2025, covered both the Novum IQ LVP and the Novum IQ syringe pump. According to FDA's notice, software anomalies can leave a blank Run screen: therapy continues, but it cannot be adjusted without interrupting it. On syringe pumps, a false motor movement system error (codes 20602 or 20603) can interrupt therapy after certain dose or rate changes below the minimum recommended flow rate on syringes of 10 mL or larger. Baxter advised customers to consider an alternate pump for high-risk medications or critically ill patients, if one is available and appropriate. In its Form 10-Q for the quarter ended 30 September 2026, Baxter states three things: FDA classified its 2025 Novum LVP corrections as Class I recalls; in July 2025 it elected to temporarily stop distributing and installing the Novum LVP in the US and Canada, except in cases of medical necessity; and some further corrections may require regulatory clearance or approval and are in the early stages of verification testing. Investors filed putative securities class actions in the Northern District of Illinois in October and December 2025. According to the 10-Q, these allege false and misleading statements and a failure to disclose material facts relating to the Novum LVP. A Baxter spokesperson told Drug Delivery Business that the company does not comment on pending litigation and intends to vigorously defend its position.[1,2,3,4,5]

Documented harm

According to FDA's notice, Baxter reported 79 serious injuries and two deaths associated with the underinfusion issue as of 27 June 2025. For the blank-screen and false-error software issues, Baxter reported three serious injuries and no deaths as of 21 July 2025. The sources do not establish causation in individual cases.

What it teaches

  • Rate increases and boluses are distinct pump operating states that warrant their own verification in testing and post-market surveillance.
  • Workarounds that require staff to reposition tubing before every bolus add bedside steps that are easy to miss under pressure; a validated alternative pump reduces reliance on them.
  • A distribution hold that lasts more than a year means hospitals may need to plan for mixed pump fleets, including the training and drug-library upkeep that two platforms require.

Sources

  1. Infusion Pump Correction: Baxter Issues Correction for Novum IQ Large Volume Pumps. U.S. Food and Drug Administration, 22 July 2025. Primary Official report · link checked 2026-11-02
  2. Infusion Pumps Correction: Baxter Issues Correction for Novum IQ Infusion Pumps. U.S. Food and Drug Administration, 15 August 2025. Primary Official report · link checked 2026-11-02
  3. Baxter International Inc. Form 10-Q for the quarterly period ended September 30, 2026. Baxter International Inc. (SEC EDGAR), October 2026. Primary Official report · link checked 2026-11-02
  4. FDA Issues Early Alert on Issue Correction for Baxter's Novum IQ Pumps. Oncology Nursing Society, August 2025. Supporting Journalism · link checked 2026-11-28
  5. Baxter faces investor lawsuit over infusion pumps. Drug Delivery Business News, 17 October 2025. Supporting Journalism · link checked 2026-11-28

Study of 2017-2025 FDA data: neurosurgical software recalls, mostly Class I, with adverse-event reports continuing after recall

A peer-reviewed analysis in Neurosurgical Focus identified 20 software-associated neurological device recalls initiated in the US from 2017 to 2025, more than half of them FDA Class I. It counted 16,333 adverse-event notifications for these devices, more than 10,000 of them reported after formal recall notification for two devices.

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

Luna G3 BPAP devices record the set pressure instead of the pressure delivered when Targeted VT is on; Class I recall

FDA classified a recall of the React Health (3B Medical) Luna G3 BPAP System, model LG3800-30VT, as Class 1. With Targeted VT enabled, the device records and transmits the configured inspiratory pressure instead of the pressure actually delivered. React Health says this could lead clinicians to adjust settings based on inaccurate data, which could delay therapy.

Tandem Mobi insulin pumps: false occlusion alarms suspend insulin delivery; Class 1 correction

Tandem Diabetes Care began a correction of Tandem Mobi insulin pumps in September 2026 because the pumps can raise false-positive occlusion alarms that suspend insulin delivery; FDA lists it as a Class 1 recall. A company spokesperson reported 230 adverse events requiring medical intervention or hospitalization as of 29 October 2026, and no deaths.

VOCSN ventilators: loose internal flow-transducer connector can cause reboots; FDA early alert and Class I recall

React Health (Ventec Life Systems) warned that the internal flow transducer cable connector in VOCSN ventilators can unseat and cause reboots and recurring System Fault 10 alarms, which could interrupt ventilation. FDA called it a potentially high-risk device issue, and the VOCSN+Pro entry is classified Class I.

All incidents · Devices

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.