According to FDA, a letter from Abiomed dated 12 August 2026 told customers that the purge flag component in the controller's purge pressure sensor assembly could fail, causing purge cassette recognition problems. These showed up as "Purge Disc Disconnected" alarms at setup and "Purge Disc not Detected" alarms during support. FDA says Abiomed recommended removing all Automated Impella Controllers in phases, with hospitals using their current inventory while waiting for service. If an alarm did not clear after the purge disc was reinserted, users were told to switch to a backup controller. Abiomed plans to replace the purge flag and retainer components. FDA classified this as a Class I recall on 24 September 2026. A Johnson & Johnson spokesperson told TCTMD that the issue had been identified in 0.2% of cases.
Abiomed sent a second removal letter on 18 September 2026 after a review of servicing records found units that needed hardware, firmware or software updates. FDA's recall record lists four problems. An RFID firmware issue may stop the controller recognising the purge cassette during setup or a cassette change, which can delay a procedure or therapy. The rotary push button may stick, which can lead to loss of control of the console. The controller may become unstable, which can cause start-up failure and an inability to drive the pump. A calculator software defect can produce negative central venous pressure values in pulmonary artery pressure calculations. The record lists 428 units distributed worldwide, with the cause still under investigation by the firm. FDA's notice says the updates were available "but have not been implemented consistently across all potentially impacted units". FDA identified this action as Class I on 9 November 2026. Separately, MedTech Dive reported a Class I correction in April 2026 for four hardware issues that could lead to a failure to boot the controller or a sudden interruption in haemodynamic support.[1,2,3,4,5]