JVN (advisory JVNVU#96989989, published 30 September 2025) and CISA (ICSMA-25-296-01, 23 October 2025) reported a NULL pointer dereference (CVE-2025-59668) in Nihon Kohden Central Monitor CNS-6201 versions 01-03 through 02-40. Processing a specially crafted UDP packet can make the device terminate abnormally; no authentication is required. CISA lists CVSS v3.1 7.5 and v4 8.7 scores, with availability as the only impact. CISA said US maintenance support for the product ended in September 2024 and that no known public exploitation had been reported to it. JVN said no patch is offered and advised users to stop using the end-of-support product.
The vendor's mitigations for sites that keep the monitors include isolating them from the Internet and untrusted systems, monitoring traffic to them, placing a firewall between the monitoring network and other networks such as the hospital information system, and keeping redundancy through a bedside monitor or telemetry. FDA's recall database lists a Class 2 correction (Z-0089-2027) of the CNS-6200 Series Central Nurse Station, model CN-6201, initiated by Nihon Kohden America on 8 September 2026, with 808 units in US commerce. The firm states that exploitation could disrupt centralised monitoring of multiple beds and the visualisation of alarms, and advises upgrading to a successor product such as the CNS-2101.[1,2,3]