Nihon Kohden CNS-6201 central monitors: unpatched denial-of-service flaw in an end-of-support product

JVN and CISA disclosed that a crafted network packet can make Nihon Kohden's CNS-6201 central monitor terminate abnormally; no patch is offered because the product is out of support. In September 2026 the firm began a Class 2 correction, saying exploitation could disrupt centralised monitoring of multiple beds and the display of alarms.

Sources checked when written 28 November 2026

TierNo exploitation or monitoring outage is documented; the advisories describe a potential loss of central monitoring. How tiers are assigned.

What happened

JVN (advisory JVNVU#96989989, published 30 September 2025) and CISA (ICSMA-25-296-01, 23 October 2025) reported a NULL pointer dereference (CVE-2025-59668) in Nihon Kohden Central Monitor CNS-6201 versions 01-03 through 02-40. Processing a specially crafted UDP packet can make the device terminate abnormally; no authentication is required. CISA lists CVSS v3.1 7.5 and v4 8.7 scores, with availability as the only impact. CISA said US maintenance support for the product ended in September 2024 and that no known public exploitation had been reported to it. JVN said no patch is offered and advised users to stop using the end-of-support product. The vendor's mitigations for sites that keep the monitors include isolating them from the Internet and untrusted systems, monitoring traffic to them, placing a firewall between the monitoring network and other networks such as the hospital information system, and keeping redundancy through a bedside monitor or telemetry. FDA's recall database lists a Class 2 correction (Z-0089-2027) of the CNS-6200 Series Central Nurse Station, model CN-6201, initiated by Nihon Kohden America on 8 September 2026, with 808 units in US commerce. The firm states that exploitation could disrupt centralised monitoring of multiple beds and the visualisation of alarms, and advises upgrading to a successor product such as the CNS-2101.[1,2,3]

Documented harm

None documented.

What it teaches

  • Once a device is out of support, network segmentation and bedside redundancy become the only controls, since no patch will come.
  • A central station that aggregates alarms from many beds is a single point of failure, and bedside alarms must stay audible on their own.
  • Asset inventories that record support end dates help hospitals plan replacement before a vulnerability is disclosed.

Sources

  1. JVNVU#96989989: NIHON KOHDEN Central Monitor CNS-6201 NULL pointer dereference. JVN (JPCERT/CC and IPA), 30 September 2025. Primary Official report · link checked 2026-11-28
  2. ICSMA-25-296-01: NIHON KOHDEN Central Monitor CNS-6201. Cybersecurity and Infrastructure Security Agency, 23 October 2025. Primary Official report · link checked 2026-11-02
  3. Class 2 Device Recall Nihon Kohden CNS-6200 Series Central Nurse Station (Z-0089-2027). U.S. Food and Drug Administration (Medical Device Recalls database), 6 October 2026. Primary Dataset · link checked 2026-11-02

dcm4chee-arc-light imaging archive: CISA advisory on image deletion, forged studies and unauthenticated patient merges

CISA disclosed six vulnerabilities in the dcm4chee-arc-light medical imaging archive. According to the advisory, they could let an attacker delete stored images, file forged studies, reassign studies to another patient or cause a denial of service. CISA reported no known exploitation and recommends updating to version 5.35.2 or later.

Philips Lumify ultrasound demands online re-registration and blocks scanning in pre-hospital settings

FDA issued an Early Alert on 9 October 2026 and on 19 October 2026 said the issue had been classified as a Class I recall. Philips had reported that its Lumify ultrasound app may stop recognising a registered transducer and block scanning until the transducer is re-registered, which requires Wi-Fi or cellular access. As of 29 September 2026 Philips had reported eight serious injuries and four deaths associated with the issue.

PathConnectivity & data → Devices

CrowdStrike Falcon content update crashes Windows hosts, including hospital systems

A faulty Rapid Response Content update to CrowdStrike's Falcon sensor crashed about 8.5 million Windows devices worldwide. Outside-in measurement found disrupted services at 759 of 2,232 US hospitals studied.

PathDevices → Connectivity & data → Human handoff

Study of 2017-2025 FDA data: neurosurgical software recalls, mostly Class I, with adverse-event reports continuing after recall

A peer-reviewed analysis in Neurosurgical Focus identified 20 software-associated neurological device recalls initiated in the US from 2017 to 2025, more than half of them FDA Class I. It counted 16,333 adverse-event notifications for these devices, more than 10,000 of them reported after formal recall notification for two devices.

All incidents · Devices · Connectivity & data

Information only, not advice. FailSystems is an aggregation and synthesis of published sources. It is not consulting, engineering, legal, regulatory or medical advice, and using it creates no professional relationship. Health systems are complex and no approach fits every organisation: anything you adopt is your own decision, at your own risk, and should be checked against the current official sources and by qualified people who know your setting. Full disclaimer.